
I-E67-TB82FF-2A February 13, 2002
2
where T is in degrees Celsius. The measured
resistance should be within the expected value by ±
5%.
Moisture intrusion behind the sensor electrode seal
can be detected with the sensor removed from the
process. Thoroughly dry and tested the sensor using
the following procedure:
1. Check the conductance between the yellow
Temperature Sensor lead and each of the other
sensor leads (i.e., green, red, white, black, and heavy
green leads). The reading must be less than 0.05 nS.
2. Check the conductance between the green Drive
lead and each of the other sensor leads (i.e., black,
white, red, and heavy green leads). The reading must
be less than 0.05 nS.
3. Check the conductance between the heavy green
lead (Shield) and each of the other sensor leads (i.e.,
blue, yellow, black, green, red, and white leads). The
reading must be less than 0.05 nS.
Transmitter Electronic Test
The TB82EC FF Transmitter can be electronically tested without requiring the use of a sensor. To perform this type
of test, use the following procedure:
1) Disconnect the sensor from the transmitter.
2) Set the transmitter temperature compensation (TC.TYPE) to MANUAL.
3) Connect the transmitter to a resistance decade box as shown in the figure below.
REAR VIEW
POWER/OUTPUT
CONNECTIONS
TB2
TB1
4
3
2
1
SIGNAL
DRIVE1
2 SENSE
3
SENSE
4DRIVE
5
RTD
RTD6
SHIELD
8
7
VOLTAGE AT POWER
TERMINALS TB1-1
AND TB1-2 MUST BE
BETWEEN 9 AND 32 VDC.
FIELDBUS
FOR
82EC
ONLY
FOR
82EC
ONLY
DECADE
RESISTANCE
BOX
Figure 35 – Sensor Simulation Setup
4) Set the decade resistance box to the values indicated in the following table:
Sensor Group Variable Resistance
(ohm)
Desired Display Value Actual Display Value
Open
0.0 µS/cm
5,000.0
100.0 µS/cm
1,000.0
500 µS/cm
500.0
1000 µS/cm
100.0 5.00 mS/cm
50.0 10.00 mS/cm
10.0 50.0 mS/cm
5.0 100.0 mS/cm
1.0 500 mS/cm
A
0.5 1000 mS/cm
Open
0.00 µS/cm
50,000
1.00 µS/cm
5,000
10.00 µS/cm
1,000
50.0 µS/cm
500
100.0 µS/cm
100
500 µS/cm
B
50
1000 µS/cm
Comentários a estes Manuais